How to Calibrate a Watlow F4T Controller

Calibrating a Watlow F4T controller is essential to maintain accurate temperature control and ensure reliable operation in various industrial applications. This comprehensive guide will walk you through the step-by-step process of How to Calibrate a Watlow F4T controller, providing you with the knowledge and confidence to achieve precise temperature control and optimize the performance of your equipment.






Step 1: Gather the Required Tools and Equipment:

When preparing to calibrate a Watlow F4T controller, it is important to gather the required tools and equipment beforehand. Here is a more detailed explanation of the tools and equipment needed for the calibration process:

For reference or to purchase each item, click the links next to each item for our recommendations

  1. Reference thermometer with known accuracy: A reference thermometer is used to measure the temperature and serves as a standard for comparison. It should have a known accuracy that is traceable to a national or international standard. Choose a high-quality thermometer with a calibration certificate to ensure accuracy. A reliable and cheap reference thermometer click here

  2. Stable heat source: A stable heat source is necessary to create a controlled environment for temperature calibration. This can be a calibration bath, dry block calibrator, or oven with stable and uniform temperature distribution. The heat source should be capable of maintaining a stable temperature throughout the calibration process.

  3. Precision multimeter: A precision multimeter is used to measure and verify the voltage output of the Watlow F4T controller. Choose a multimeter with high accuracy and resolution to ensure precise readings. It should be capable of measuring both DC voltage and resistance.

  4. Calibration software: Depending on the specific requirements and features of the Watlow F4T controller, you may need calibration software that is compatible with the controller. This software allows you to communicate with the controller, configure calibration parameters, and log calibration data. Ensure that the software supports the necessary communication protocols and functions required for the calibration process.

Other additional tools and equipment that may be needed include:

  • Calibration standards: Depending on the specific temperature range and accuracy requirements, you may need additional calibration standards such as calibrated resistors, thermocouples, or RTDs (Resistance Temperature Detectors).

  • Connection cables and adapters: Ensure that you have the appropriate cables and adapters to connect the reference thermometer, precision multimeter, and any other equipment to the Watlow F4T controller. Check the connector types and compatibility to ensure proper connections.

  • Calibration certificates and documentation: Keep all relevant calibration certificates, user manuals, and documentation for the equipment used in the calibration process. This documentation is important for traceability and ensuring compliance with quality standards.



Step 2: Understand the F4T Calibration Procedure

Familiarize yourself with the calibration procedure outlined in the Watlow F4T controller's user manual. This procedure may vary depending on the specific model and firmware version. It is essential to follow the manufacturer's guidelines to ensure accurate calibration and prevent any potential issues.


Dont have the user manual?

Click here to download the Watlow F4T user manual



Step 3: Prepare the F4T for Calibration Setup:


Create a stable and controlled environment for the calibration process. Ensure that the ambient temperature remains constant throughout the calibration. Connect the reference thermometer to the Watlow F4T controller and the precision multimeter for accurate temperature and voltage readings.


  • Create a stable and controlled environment: Find a location where the ambient temperature remains constant during the calibration process. This helps eliminate external factors that could affect the calibration results. Avoid areas with drafts or temperature fluctuations.

  • Connect the reference thermometer: Use a reliable and accurate reference thermometer to measure the temperature. Connect the reference thermometer to the Watlow F4T controller using the appropriate interface or connection method recommended by the manufacturer. Ensure the connection is secure and stable.

  • Connect the precision multimeter: To measure the voltage output of the controller, connect a precision multimeter to the appropriate terminals on the Watlow F4T controller. Follow the manufacturer's instructions for proper connection. It is essential to use a precise multimeter to obtain accurate readings.

  • Ensure proper calibration settings: Before starting the calibration process, verify that the calibration settings on the Watlow F4T controller are correctly configured. This includes selecting the calibration mode, temperature units, and any other relevant settings that may affect the calibration procedure.

  • Monitor temperature and voltage readings: Throughout the calibration process, closely monitor the temperature readings displayed on both the reference thermometer and the Watlow F4T controller. Compare the readings to ensure consistency and accuracy. Similarly, observe the voltage readings on the precision multimeter for any deviations or inconsistencies.


Step 4: Adjust the Calibration Settings:

Access the calibration settings menu of the Watlow F4T controller through the user interface or the calibration software. Follow the provided instructions to enter the calibration mode and adjust the necessary parameters. Typically, these settings include offset and gain adjustments to align the controller readings with the reference thermometer.


To calibrate the Watlow F4T controller accurately, you need to adjust the calibration settings. Here is a more detailed explanation of the steps involved:

  • Access the calibration settings menu: Depending on the specific model and firmware version of the Watlow F4T controller, you can access the calibration settings menu through the user interface (if available) or the calibration software. Refer to the user manual or documentation provided by Watlow for instructions on how to access this menu.

  • Enter the calibration mode: Once you have accessed the calibration settings menu, follow the provided instructions to enter the calibration mode. This mode allows you to make adjustments to the necessary calibration parameters.

  • Adjust the offset and gain: The calibration settings typically include offset and gain adjustments. The offset adjustment allows you to align the controller readings with the reference thermometer at a specific temperature point. The gain adjustment helps in adjusting the slope or sensitivity of the controller's temperature readings.

  • Align with the reference thermometer: Use the reference thermometer connected to the controller and the precision multimeter to monitor the temperature and voltage readings during the calibration process. Compare the readings from the controller with the readings from the reference thermometer.

  • Make necessary adjustments: Based on the comparison between the controller readings and the reference thermometer readings, make adjustments to the offset and gain settings as needed. The goal is to minimize any deviation or difference between the two readings and ensure accurate temperature control.

  • Verify and fine-tune: After making the initial adjustments, recheck the controller readings and compare them with the reference thermometer readings. Fine-tune the calibration settings if necessary to achieve closer alignment and better accuracy.

  • Save the calibration settings: Once you are satisfied with the calibration adjustments, save the calibration settings in the Watlow F4T controller. Follow the instructions provided in the calibration settings menu or software to save the changes effectively.


Step 5: F4T Zero Calibration:

Begin the calibration process by performing the zero calibration. This step involves setting the controller to a known zero point temperature, ensuring that the controller outputs zero when the temperature is at the reference point. Follow the user manual instructions to execute this calibration correctly.

Performing the zero calibration is an important step in calibrating the Watlow F4T controller. Here's a more detailed explanation of how to carry out this calibration process:

  • Understand the purpose: Zero calibration establishes the baseline or reference point for the controller's temperature readings. It ensures that when the temperature is at this reference point, the controller outputs zero, indicating no deviation from the baseline.

  • Refer to the user manual: Consult the user manual or documentation provided by Watlow for specific instructions on how to perform the zero calibration for the F4T controller. The manual will provide detailed step-by-step guidance tailored to your specific model and firmware version.

  • Access the calibration settings menu: To initiate the zero calibration, access the calibration settings menu on the Watlow F4T controller. This can be done through the user interface (if available) or the calibration software.

  • Enter the zero calibration mode: Follow the instructions provided in the user manual to enter the zero calibration mode. This mode enables you to set the controller to the known zero point temperature.

  • Set the reference temperature: Once in the zero calibration mode, adjust the temperature settings to the known zero point temperature. This could be room temperature (if ambient temperature is stable) or a specific temperature point defined by the calibration standards or requirements.

  • Verify the output: After setting the temperature, observe the controller's output. It should read zero or very close to zero at the reference temperature. This indicates that the controller is calibrated to recognize and compensate for any deviations from the baseline at the zero point.

  • Save the calibration settings: If the zero calibration is successful, save the calibration settings in the Watlow F4T controller. Follow the instructions provided in the calibration settings menu or software to save the changes.

  • Perform additional verifications: It is good practice to perform additional checks or verifications after completing the zero calibration. This can include comparing the controller's output at other known temperatures with the reference thermometer to ensure accurate temperature readings throughout the range.

Remember to follow the manufacturer's instructions and guidelines for performing the zero calibration accurately. It is also advisable to maintain a record of the zero calibration procedure, including any calibration certificates or documentation, for future reference and traceability.



Step 6: F4T Span Calibration:

Performing the span calibration is an essential step in calibrating the Watlow F4T controller. This calibration process establishes the relationship between the input temperature and the output control signal of the controller. Here's a more detailed explanation of how to perform the span calibration:

  • Understand the purpose: Span calibration ensures that the controller's output control signal corresponds accurately to the input temperature. By applying a stable and known heat source, you can adjust the calibration settings to align the controller's readings with the reference thermometer's readings.

  • Gather the necessary equipment: Before starting the span calibration, ensure you have a stable and known heat source available. This could be a calibrated temperature source or a reference oven. Additionally, you will need the reference thermometer, the precision multimeter, and the appropriate calibration software compatible with the Watlow F4T controller.

  • Access the calibration settings menu: Enter the calibration settings menu of the Watlow F4T controller either through the user interface or the calibration software. Follow the provided instructions in the user manual to access the appropriate settings for the span calibration.

  • Enter the span calibration mode: Follow the instructions provided in the user manual or the calibration software to enter the span calibration mode. This mode allows you to adjust the calibration settings to match the reference thermometer's readings.

  • Apply the stable heat source: Apply the stable and known heat source to the controller. Ensure that the temperature remains constant throughout the calibration process. It is recommended to wait until the temperature stabilizes before proceeding.

  • Compare the readings: Using the reference thermometer and the precision multimeter, measure and record the temperature readings from both the Watlow F4T controller and the reference thermometer. Compare the readings to identify any deviations or discrepancies.

  • Adjust the calibration settings: Based on the comparison of the controller's readings with the reference thermometer, make the necessary adjustments to the calibration settings. Typically, these settings include offset and gain adjustments to align the controller's output control signal with the reference thermometer's readings.

  • Verify and fine-tune: After adjusting the calibration settings, recheck the temperature readings from both the controller and the reference thermometer. Fine-tune the calibration settings as needed to achieve a closer match between the two readings.

  • Save the calibration settings: Once the span calibration is complete and the controller's output control signal is aligned with the reference thermometer's readings, save the calibration settings in the Watlow F4T controller. Follow the instructions provided in the calibration settings menu or software to save the changes.

  • Perform additional verifications: It is good practice to perform additional checks or verifications after completing the span calibration. This can include applying different temperature points and comparing the controller's output control signal with the reference thermometer's readings to ensure accurate temperature control across the desired range.

Always follow the manufacturer's instructions, guidelines, and best practices when performing the span calibration. Maintain proper documentation, including calibration certificates and records, for future reference and traceability.


Step 7: Verify and Document Calibration Results:

After performing the calibration process on the Watlow F4T controller, it is crucial to verify and document the calibration results to ensure the accuracy of the controller's temperature readings and control signal output. Here's a more detailed explanation of the verification and documentation steps:

  • Verify temperature readings: Using the reference thermometer, compare the temperature readings displayed on the Watlow F4T controller with the readings from the reference thermometer. Check for any discrepancies or deviations between the two measurements.

  • Measure control signal output: Using the precision multimeter, measure and record the control signal output of the Watlow F4T controller. This will help confirm that the controller is producing the expected control signal based on the input temperature.

  • Compare and analyze results: Compare the temperature readings from the Watlow F4T controller with the reference thermometer and the control signal output with the expected values. Analyze the results to identify any differences or inconsistencies.

  • Adjustments, if necessary: If there are any significant discrepancies or deviations found during the verification process, revisit the calibration settings and make any additional adjustments as needed. Repeat the verification process to ensure the accuracy of the calibration.

  • Document calibration results: It is essential to maintain proper documentation of the calibration process and its results. Record the initial and final calibration settings, temperature readings, control signal outputs, and any adjustments made during the calibration process. This documentation serves as a reference for future calibration activities and provides traceability.

  • Calibration certificate: If applicable, generate a calibration certificate that includes all relevant information about the calibration, such as date, equipment used, calibration procedure, calibration standards, and the calibration results. This certificate serves as official documentation and proof of the calibration performed.

  • Retain records: Store the calibration records, including the documentation and calibration certificate, in a secure and organized manner. This allows for easy retrieval and reference in the future, such as during audits or when evaluating the performance of the controller over time.

By verifying and documenting the calibration results, you ensure the accuracy and reliability of the Watlow F4T controller. It also enables you to demonstrate compliance with quality standards and provides a historical record of the controller's calibration for future reference.


Calibrating a Watlow F4T controller is a critical task to ensure accurate temperature control and reliable operation of your industrial equipment. By following the step-by-step instructions outlined in this guide, you can calibrate your Watlow F4T controller with confidence, achieving precise temperature control and optimizing the performance of your system. Regular calibration will help maintain accuracy and enhance the overall efficiency of your processes.